8. Troubleshooting
Symptom
Sample Chambers are not
filling
100
Possible Cause
Reagent Tube not properly connected.
Reagent Tube kinked.
Bottle filter obstructed.
Tube not properly connected on the Module
Manifold.
The inlet/outlet opening in the Sample
Chamber may be obstructed. See image
below.
Reagent Bottle is empty.
The Reagent Pump is not pumping reagent,
see image below.
Remedies
Check the tube connections on the
Reagent Manifold and Reagent Bottle
cap.
Clean the bottle filter with 70% ethanol.
Replace Filter Assembly.
Check the tube connections on the
Reagent Manifold. See
Module
Clean the system; see
7.3 - Line
If the problem persists, call Technical
Support or your distributor.
Refill the Reagent Bottle with the
appropriate reagent.
Check if the rollers are turning on the left
pump located on the front of the
instrument.
If not, call Technical Support or your
distributor.
ThermoBrite Elite User Manual
3855-7500-501 B02
Manifold .
Wash .