Symptom
Sample Chambers are not
emptying
ThermoBrite Elite User Manual
3855-7500-501 B02
Possible Cause
Tube not properly connected on the Module
Manifold.
The Waste Pump is not pumping waste liquid,
see image below.
The inlet/outlet opening in the Sample
Chamber may be obstructed. See image
below.
Clogged Manifold Filters
8. Troubleshooting
Remedies
Check the tube connections on the
Reagent Manifold. See
Module
Check if the rollers are turning on the
right pump located on the front of the
instrument.
If not, call Technical Support or your
distributor.
Clean the system; see
7.3 - Line
If the problem persists, call Technical
Support or your distributor.
Replace the Manifold Filters; see
7.6.1 - Replace the Module Manifold
Filters.
Manifold .
Wash .
101