5. Click Update to validate the entry. The Reagent Step screen is displayed.
New steps are added immediately after the currently highlighted step.
Pause
Use this option to add a specific instruction and to pause the protocol run for a manual user action.
1. Select the step to precede the Pause step, and then click the Pause button.
2. Enter a Comment. During Pause, the Thermal Modules will hold temperature until the user clicks Resume. A
Pause step may be useful post hybridization incubation to allow the user to return for post wash when ready.
3. Click Update to validate the entry. The Reagent Step screen is displayed.
ThermoBrite Elite User Manual
3855-7500-501 B02
Figure 6-31: Protocol Reagent Step—Prompt
6. Using the ThermoBrite Elite Software
75