8.2.20 The protocol has failed

Causes
A problem caused the protocol to fail.
8.2.21 This protocol contains errors and it cannot be
run
Causes
The selected protocol contains errors in the configuration. Verify the configuration steps and sub-steps.

8.2.22 Unhandled exception

Causes
An internal component is not responding to a command.

8.2.23 USB Communication interruption

Causes
The Netbook lost connection with the ThermoBrite Elite.
ThermoBrite Elite User Manual
3855-7500-501 B02
Remedies
Restart the Netbook and instrument.
Restart the protocol.
If the problem persists, contact Technical Support or your
distributor.
Remedies
Restart the Netbook.
If the problem persists, contact Technical Support or your
distributor.
Remedies
Restart the Netbook and instrument.
If the problem persists, contact Technical Support or your
distributor.
Remedies
Unplug the USB from the Netbook and then plug it in again.
Restart the Netbook and instrument.
If the problem persists, contact Technical Support or your
distributor.