7. Maintenance

7.4.1.3 Clean the Pepsin Bottle

Items required: Acceptable disinfectants are 70% or higher concentration alcohol, deionized water, lint-free tissue, and
appropriate personal protective equipment.
1. Ensure the instrument is not in operation.
2. Remove the cap and empty the Pepsin Bottle.  Dispose of the reagent in accordance with approved procedures at
your facility.
3. Wash the bottles with 70% alcohol.
4. Rinse thoroughly with deionized water.
5. Allow the containers to dry before refilling with fresh reagent and returning to the instrument.

7.4.1.4 Clean the Pepsin Line

After each run, perform a line wash using 70% or higher alcohol on the Pepsin line. See
7.4.2

Monthly

Perform the following cleaning procedures monthly.

7.4.2.1 Clean Instrument Surfaces

Items required: Mild detergent, lint-free tissue, distilled water, and appropriate personal protective equipment.
1. Clean the outside surfaces of the ThermoBrite Elite Unit with a water-dampened lint-free tissue and mild detergent
to remove any deposits.
2. Wipe again using distilled water.
3. Wipe dry.

7.4.2.2 Clean Slide Carriers

Items required: Acceptable disinfectants are 70% or higher concentration alcohol, lint-free tissue, distilled water, and
appropriate personal protective equipment.
1. Remove any Slide Carriers from inside the instrument.
2. Soak the Slide Carrier in 70% or higher concentration alcohol for at least 5 mins. Do a visual check to ensure it is
clean.
3. Rinse thoroughly using distilled water.
4. Wipe dry with a lint-free tissue.
90
7.3.1 - Run a line wash
ThermoBrite Elite User Manual
3855-7500-501 B02