As Needed Maintenance; Replace The Module Manifold Filters - Leica THERMOBRITE ELITE User Manual

Table of Contents

Advertisement

7.6

As Needed Maintenance

7.6.1

Replace the Module Manifold Filters

The Module Manifold is located on the front of the instrument between the peristaltic pumps. If the instrument has not
been maintained adequately, you might need to replace the Module Manifold Filters in between services. When running
protocols, if you see multiple messages relating to either slow fill or critical fill errors for a chamber, you might need to
replace the Module Manifold Filters.
A single error relating to problems with the fill does not necessarily indicate that the Module Manifold Filters
need to be replaced. Wait until there are multiple errors for the same chamber.
Items required: Manifold Filter Kit, Item# 3800-007743-001, paper towel, and appropriate personal protective equipment.
1. Turn the main power switch located in the back panel to the Off (O) position. Make sure the power cable to the
ThermoBrite Elite Unit is unplugged.
2. Place paper towel underneath the Module Manifold to catch any possible fluid leak (see image below).
3. Using a flat screwdriver and turning counterclockwise, remove each Manifold Filter.
4. Discard the old Manifold Filters according to applicable regulations.
ThermoBrite Elite User Manual
3855-7500-501 B02
Figure 7-7: Module Manifold
Dispose of contaminated material in compliance with applicable legal regulations. If unsure of
the applicable legal requirements, contact the local authorities for information.
7. Maintenance
93