LEICA TP1020 – Automatic Tissue Processor
The manual processing mode
The carousel may not be rotated manually!
!
Severe damage will result from doing so!
Step 1: Inserting the tissue basket
Fill the tissue cassettes or capsules into the tissue basket.
Lift the carousel and rotate the carousel so the basket holder (1) is
located above the starting container for basket 1
Hook the tissue basket (2) into the basket holder as shown.
Lower the basket into the starting container.
Pressing the corresponding button.
To lift the carousel, press ARROW UP.
To lower the carousel, press ARROW DOWN
To move on to the next station, press CIRCLE ARROW

Advertisement

loading

Summary of Contents for Leica TP1020

  • Page 1 LEICA TP1020 – Automatic Tissue Processor The manual processing mode Pressing the corresponding button. The carousel may not be rotated manually! Severe damage will result from doing so! To lift the carousel, press ARROW UP. To lower the carousel, press ARROW DOWN...
  • Page 2 Step 2: Starting a program The program on display is always the last one that has been in use. The last program is often the number To start a program, press start For to change the program: Select the desired program number by pressing PLUS or MINUS To start the selected program immediately, press START again Step 3: Remove the specimens To switch off the function up and down, press THREE ARROW...