8.2.15 No ThermoBrite Elite found
Causes
The Netbook cannot connect to the ThermoBrite Elite.
8.2.16 Reagent valve failed to switch to reagent line #
Causes
The reagent valve is not responding correctly.
8.2.17 Self-test failed to communicate with
Instrument
Causes
The Netbook cannot connect to the ThermoBrite Elite.
ThermoBrite Elite User Manual
3855-7500-501 B02
Remedies
Check the USB connections between the instrument and
the Netbook.
Restart the Netbook and instrument.
If the problem persists, contact Technical Support or your
distributor.
Remedies
Restart the Netbook and instrument.
If the problem persists, contact Technical Support or your
distributor.
Remedies
Unplug the USB from the Netbook and then plug it back in.
Restart the Netbook and instrument.
If the problem persists, contact Technical Support or your
distributor.
8. Troubleshooting