7. Maintenance
7.3.1

Run a line wash

Items required: 70% ethanol (or reagent grade alcohol or deionized water), paper towel, and appropriate personal
protective equipment.
For Clearene and line washes, use 100% ethanol.
1. Insert then twist to lock up to 3 Reagent Line fittings onto the cap of the ThermoBrite Elite Wash Bottle (filled with
deionized water, 70% ethanol, or 100% ethanol for Clearene lines).
2. From the Main screen, click the Line Wash button.
86
Figure 7-1: Insert Reagent Line Fittings onto Wash Bottle
Figure 7-2: Select Line Wash
ThermoBrite Elite User Manual