Sensors Indicate A Critical Flow Error; Sensors Indicate Reduced Flow Rate Warning(S) - Leica THERMOBRITE ELITE User Manual

Table of Contents

Advertisement

8. Troubleshooting

8.2.18 Sensors indicate a critical flow error

Causes
Sensors indicate a critical flow problem while filling.
Causes:
empty reagent bottle
l
disconnected reagent tube
l
blocked reagent bottle filter
l
worn Peristaltic Tube
l
Sensors indicate a critical flow problem while draining.
Causes:
obstructed Chamber Port
l
blocked Manifold Filter
l
worn Peristaltic Tube
l
full Waste Bottles
l

8.2.19 Sensors indicate reduced flow rate warning(s)

Causes
Sensors indicate a reduced flow problem while filling.
Causes:
empty reagent bottle
l
disconnected reagent tube
l
blocked reagent bottle filter
l
worn Peristaltic Tube
l
Sensors indicate a reduced flow problem while draining.
Causes:
obstructed Chamber Port
l
blocked Manifold Filter
l
worn Peristaltic Tube
l
full Waste Bottles
l
106
Remedies
Restart the Netbook and instrument.
Perform a Line Wash, see
l
Refill any empty Reagent Bottle
l
Check the Reagent Tubes connections to the
l
Reagent Bottles and the Reagent Manifold.
Replace the Peristaltic Tubes; see
l
the Peristaltic Pump
If the problem persists, contact Technical Support or your
distributor.
Remedies
Restart the Netbook and instrument.
Perform a Line Wash, see
l
Refill any empty Reagent Bottle.
l
Check the Reagent Tubes connections to the
l
Reagent Bottles and the Reagent Manifold.
Replace the Peristaltic Tubes; see
l
the Peristaltic Pump
If the problem persists, contact Technical Support or your
distributor.
7.3.1 - Run a line wash
7.6.2 - Replace
Tubes.
7.3 - Line
Wash .
7.6.2 - Replace
Tubes.
ThermoBrite Elite User Manual
3855-7500-501 B02