Reagent Rock
This option is used to define the cycle rate at which the ThermoBrite Elite will rock back and forth. The default is Medium
speed – 12 cycles/minute.
We recommend not using Fast as it might damage the sample.
1. Double-click Rock.
2. Select the appropriate option:
Stop - the instrument will not rock
l
Slow - 9 cycles/minute
l
Medium - 12 cycles/minute
l
Fast - 20 cycles/minute
l
Selecting Stop will result in poor temperature control. Reagent steps that involve thermal control
should not be set to Stop.
3. Click Update to validate the entry. The Reagent Step screen is displayed.
ThermoBrite Elite User Manual
3855-7500-501 B02
Figure 6-27: Protocol Reagent Step—Rock
6. Using the ThermoBrite Elite Software
71