Items required: Peristaltic Pump Tube Kit, Item# 3800-007742-001, paper towel, and appropriate personal protective
equipment.
1. Turn the main power switch located in the back panel to the Off (0) position. Make sure the power cable to the
ThermoBrite Elite Unit is unplugged.
2. Place some paper towels underneath the peristaltic pump to catch any possible fluid leak, see image below.
3. Lift the Peristaltic Pump Cover to access the pump. This raises the top part of the pump, giving full access to the
tube, see
Figure 7-10 - Remove Plug from Exit
4. Remove the gray plug from the exit port by pulling upward on the plug and the tube, see image below.
5. Repeat step 4 for the inlet port until the tube and the gray connectors are free.
ThermoBrite Elite User Manual
3855-7500-501 B02
Figure 7-9: Peristaltic Pump
Port.
Figure 7-10: Remove Plug from Exit Port