4.21 Replace power supply fuses

BOND-PRIME User Manual, 91.7500.501 A07
Copyright © 2023 Leica Biosystems Melbourne Pty Ltd
1.
Ensure you have the correct fuses (F1, F2).
The fuses are located at the rear of the processing module.
You may need to move the processing module
for better access to the fuses and mains-
power connector on the rear-panel.
2.
Power down the processing module.
a.
Press the standby power button.
A pop up window indicates that it is safe to shut down
the processing module.
b.
Tap Close.
3.
Power off the processing module.
a.
Open the Reservoir Cabinet door.
b.
Turn the AC power switch counterclockwise.
c.
Close the Reservoir Cabinet door.
WARNING: Take care not to trip over the
Reservoir Cabinet door when it is in the open
position.