9.3.2 Replacing the fuses

Turn the instrument off and disconnect the power plug before replacing the fuses.
Use only the fuse types specified in Chapter
Failure to comply with this can cause severe damage to the instrument and its surroundings
and a risk of fatal injury!
Fuse
Protection
F1
Display
F2
Coarse feed
F3
Processor board supply
F4
Peltier element
62
"3. Technical
On the back of the instrument there is a fuse box with 4 fuses:
• Unscrew the fuse cap with a screwdriver.
• Remove both fuse cap and fuse.
• Insert the defined fuse into the sleeve, then use the screw-
Fig. 56
driver to screw the sleeve back in as far as it will go.
Type
T 0.25 A
T 1.6 A
T 1.0 A
T 6.25 A
Data".