9.3.2 Replacing the fuses

Warning
• Turn the instrument off and disconnect the power plug before replacing the fuses!
• Use only the fuse types specified in (→ P. 17 – 3. Technical Data).
• Failure to comply with this can cause severe damage to the instrument and its surroundings as well
as endangering life and limb!
On the back of the instrument there is a fuse board with 4 fuses (→ "Fig. 58"):
1. Unscrew the sleeve of the defective fuse using the screwdriver.
2. Remove the sleeve and fuse.
3. Insert the defined fuse into the sleeve, then use the screwdriver to screw the sleeve back in as
far as it will go.
Fig. 58
Fuse
Function/protection
F1
Empty
F2
Coarse feed
F3
Processor board power supply
F4
Heaters
Leica CM1520
Cleaning, Disinfection, Maintenance
Type
T 1.6 A
T 1.0 A
T 6.25 A