Objet500 User Guide
If the UV level measured is not acceptable, the wizard calibrates the UV lamp by 
adjusting the current supplied to it, and then tests it again.
When all UV tests have finished, the results are displayed, showing the condition 
of the lamps after calibration (see Figure 6-102 below).
Figure 6-102    Results and condition of UV lamps after calibration
After examining the UV calibration results:
1.  Click Next.
2.  Remove the UV sensor and close the printer cover. After confirming the items 
DOC-07027 Rev. I
listed in the wizard screen, click Next.
If the results are not acceptable for quality printing, this is indicated in the final 
wizard screen.
Figure 6-103    Final UV Calibration Wizard screen
6 Operating and Maintaining the Printer
6-72