About Decade Display
⚫
Each item is defined as follows:
SPAN
: Span lower-limit value (Span L)
L
SPAN
: Span upper-limit value (Span R)
R
SCAL
: Scaling lower-limit value (Scale L)
L
SCAL
: Scaling upper-limit value (Scale R)
R
IN
: Input voltage
OUT
: Output (Scaling value)
⚫
Decade display abides by the following relational expression.
LG SCAL
OUT1= IN×
SPAN
OUT = 10
OUT1
LG SCAL
:Log
L
LG SCAL
:Log
R
[Note]
Decade output is used for display and print. It is not reflected on a printing position.
Example) When makes setting in previous page, the display is as follow table.
Input voltage (V)
Display
XX: Mantissa section (1.0 to 9.9)
YY: Exponent section (-19 to 19)
Up to 5 decades.
(Scaling upper limit)-(Scaling lower limit) is 1.0E5 or less.
-LG SCAL
R
L
+
- SPAN
R
L
(SCAL
)
10
L
(SCAL
)
10
R
0.0
1.0
1.0E0
1.0E1
―53―
XXEYY
×LGSCAL
- SPAN
SPAN
R
L
- SPAN
SPAN
R
2.5
3.0
3.2E2
1.0E3
×LGSCAL
L
R
L
5.0
1.0E5
HXPRM10mnCT001E