TDK-Lambda G+1U 5kW User Manual page 146

Built-in lan ( 1.5), usb, rs-232 & rs-485 interfaces built-in remote isolated analog program/monitor/control interface optional interface: is420, ieee488.2 scpi (gpib), modbus tcp or ethercat
Table of Contents

Advertisement

SYSTem:DATE?
Query
Returns the date of last calibration. Format yyyy/mm/dd
Return
<SRD>
Example
2017/12/17
SYSTem:ERRor:ENABle
Function
Enables log of last 10 Error messages.
SYSTem:ERRor?
Query
Returns an error number and a corresponding error message stored in an
error queue. Works as FIFO. When no error exists 0,"No error" is returned.
If Buffer reaches above 10 messages, the command returns -350,"Queue
Overflow".
Return
<SRD>
Example
0,"No error"
SYSTem:FRST <DSC>
Function
Restores factory default parameters and sets communication interface
according to <DSC> parameter.
Refer to Table 2-6 & Table 2-7.
Parameters
1|2|3|4|5, USB|RS232|RS485|LAN|OPTion
SYSTem:FRST without a parameter sets USB communication interface.
1. Factory Reset does not affect advanced parallel configuration acknowledgment (does not affect
power supply configuration (Single, Master, or Slave roles do not change).
2. Execution time for this command is ~250mS.
3. <DSC> Parameter is available for firmware revision 02.106 and above.
SYSTem:FIRMware[:VERSion]?
Query
Returns firmware versions string: Interface, Master Control, Display, and
Option (if available).
Return
<SRD> xx.xxx,xx.xx,x.x,xx.xxx
Option firmware revision (last xx.xxx field) is seen if the query is sent through optional
communication interface.
SYSTem:PANel:LOCK?
Query
Returns the front panel unlocked (0)/locked (1) state.
Return
<Bool> 0|1
NOTES
NOTE
131