6.7 Adjustment

As the acceleration value due to gravity is not the same at every location on earth,
each display unit with connected weighing plate must be coordinated - in compliance
with the underlying physical weighing principle - to the existing acceleration due to
gravity at its place of location (only if the weighing system has not already been
adjusted to the location in the factory). This adjustment process must be carried out
for the first commissioning, after each change of location as well as in case of
fluctuating environment temperature. To receive accurate measuring values it is also
recommended to adjust the display unit periodically in weighing operation.
Observe stable environmental conditions. Stabilisation requires a certain
warm-up time.

6.7.1 Manual internal adjustment by pressing button

 In weighing mode keep
 The motor noise of the loading system for the internal adjustment weight can be
heard, the internal adjustment is started.
After successful adjustment „PASS" will be displayed. The balance returns
automatically into weighing mode.