Caution On Installation Location - Fuji Electric FSZ Instruction Manual

Integral ultrasonic flowmeter s-flow
Table of Contents

Advertisement

CAUTION ON INSTALLATION LOCATION

Places where the ambient temperature and humidity are -15 to +60°C and 95% RH or less.
(1)
Indoor or outdoor locations where there is no exposure to direct sunlight or wind and rain.
(2)
A place that provides enough space for periodic inspection and wiring work.
(3)
A place not subjected to radiated heat from a heating furnace, etc.
(4)
A place not subjected to corrosive atmosphere.
(5)
A place not to be submerged.
(6)
A place free from excessive vibration, dust, dirt, and moisture.
(7)
A place remote from electrical devices (motor, transformer, etc.) which generate
(8)
electromagnetic induction noise, electrostatic noise, etc.
A place not subjected to excessive fluid pulsation such as pump discharge side.
(9)
A place that provides enough place for the length of the straight pipe.
(10)
Altitude: up to 2000 m
(11)
- v -