3.12.2. Checking the status output and total pulse operation
Description
●
Status simulated output (DO1, DO2)
This function checks the status output.
Setting ON: Sets the contact to short-circuit.
●
Total pulse simulated output (DO1 only)
This function checks the total pulse output.
You can check the output by outputting one pulse per second.
Setting STOP: Stops simulated pulses.
Note 1: The output pulse width corresponds to the currently selected pulse width.
●
This operation produces the same output for DO1 and DO2 at the same time.
●
Before activating it, confirm in advance that you want to change the DO output.
Operation (example)
Display and keystrokes
1
Measurement display
Press and hold the ○ key
2
○ Key
3
○ Key
4
○ Key
5
○ Key
6
○ Key
7
○ Key
8
Measurement display
OFF: Opens the contact.
GO: Outputs simulated pulses.
Press and hold the ○ key to go to Display 2.
Press the ▽ key four times and select "SET MENT".
Press and hold the ○ key to go to Display 3.
Press the ○ key to go to Display 4.
Displays "MENT DO".
Press and hold the ○ key to go to Display 5.
Status simulated output: OFF,
Use the △▽ keys and set OFF or ON to check the contact output.
Press the ○ key to go to Display 6.
Total pulse simulated output: STOP,
Use the △▽ keys and set GO to output simulated pulses.
Use the △▽ keys and set STOP to stop simulated pulses.
Press the ○ key to go to Display 7.
Press the ○ key to return to the measurement display.