Analog Output / Alarm Record - Fuji Electric FE ZSS Instruction Manual

Cross stack laser gas analyzer
Table of Contents

Advertisement

Note
Set the value so that H-Limit is larger than L-Limit, and that (H-Limit – L-Limit) is larger than hysteresis
width.

6.3.3 Analog output / alarm record

Set the recording range of the external alarm output such as analog output or the alarm record.
Setting range can be selected from "H-Limit", "L-Limit" and "H/L Limit". When there are two ranges, the
setting range can be selected from the 6 types of Range1 and Range2.
Note
When "OFF" is set to the ON/OFF setting of the alarm value, "Analog Output / Alarm Record" cannot be
set. Select "ON" again. The alarm output which is outside the range of High/Low limit cannot be per-
formed for 10 minutes after turning on the power.
Alarm Setting
H - Limit
L-Limit
Hysteresis
Analog Output
Alarm Record
Alarm Setting
H - Limit
L-Limit
Hysteresis
Analog Output
Alarm Record
50.00
Range1
Range2
30.00
ON
10.00
Range1
00.00
Range2
15
Range1 H- Limit
(
)
50.00
Range1
ppm
30.00
ppm
Range2
ON
10.00
ppm
Range1
00.00
ppm
Range2
15
%FS
Range1 H- Limit
(
)
NT
ppm
ppm
ppm
ppm
Point the
to "Analog Output / Alarm Record"
%FS
by the
key and the
the
key.
Select output range by the
key, and press the
- 42 -
key, and press
key and the
NT
key.
INZ-TN1ZSS-E