Chapter 7 Configuration and Setup
ANALYSIS SETUP PAGE 1
Pressing 5 (Analysis) from the Setup Menu causes the display to ask for the passcode if
the analyzer is configured to be passcode-protected. If the passcode is correct, the first of
four Analysis Setup Menus is displayed (see Analysis Setup Menu Page 1, above). The
current application setup summary is displayed on the left and the menu of selections on
the right. In the summary, the Logging mode is described as: Off, Single sample (single
sample per logging trigger), Continuous (continuous sampling per logging trigger), or
Custom (customized logging configuration); the Alarm state is indicated as High, Low, or
STEL; and the compensations indicated as Temperature, Pressure, Reference, and
Humidity.
To move to one of the other three Analysis Setup menus, press the up or down arrow
keys one or more times. Press
In the summary shown in the Analysis Setup Menu Page 1, the Logging mode is shown
as Single Sample and the Logging interval is set at 10 seconds (the unit counts down 10
seconds after the trigger). The Alarm setpoint is triggered on the High level. All
Compensations are turned off.
Logging
To change the logging characteristics, press
Page 1. The Log Menu is displayed.
The summary of the present log configuration is shown on the left. The selections at the
bottom of the display indicate the type of log: Mx = maximum, Min = minimum, Avg =
average, Abs = absorbance, and Tmp = temperature. In the example given in Log Menu,
above, the log will only store the Maximum value concentration in the 5-second
countdown interval.
to return to the Setup Menu.
ESC
┌────────────────────────────────────────┐
│╒══════LOG SUMMARY══════╕
││
Mode: Single Sample│1=Log mode
││Interval: 5 seconds
││
Start: Manual
││
││Store:_OnAlrmOnly _STEL│5=Other parms
││√Mx _Min _Avg _Abs _Tmp│
│└───────────────────────┘
└────────────────────────────────────────┘
Log Menu
(Logging) from the Analysis Setup Menu
1
LOG MENU
│2=Log interval │
│3=Log start
│4=Log store