Chapter 7 Configuration and Setup
Menu Item
1 = Zero save
2 = Arrow keys
3 = Log Warning
4 = Fault action
5 = Alarm
action
6 = Purge flag
Selections
Selects the zero save mode: 1 = Auto and 2 = Manual. In Auto Mode,
when a zero has been performed, a valid zero is automatically saved
and the next prompt displayed. In Manual Mode, when a zero has
been performed, there is a prompt that the zero is valid and instructed
to Save or Repeat Zero (see Chapter 5, Save Zero Display screen).
Data View enables viewing the data, described in Chapter 5, "To
Begin Analysis". by using the arrow keys. Pressing 1 = Data View,
Pressing 2 = Locations. If this is chosen, the various data can still be
viewed by using the 6 and 9 keys as arrow keys.
Automatically logs if and when an alarm or fault or both occurs. Press
(Off) to turn off this feature,
1
(On alarm only) to log at an alarm, and
fault.
Specifies the type of notification received if a fault occurs. Press
(None) if only a symbol should be displayed,
signal is also desired, and
in addition to the symbol and beep.
Specifies the type of notification received, provide a concentration
alarm occurs. Selections are the same as "Fault Action."
Enables receiving a "P" display indicator if a purge is not completed.
Press
(Off) to turn off this feature and
1
(On) to log at an alarm or a fault,
2
(On fault only) to log at a
4
(Error Screen) if an error screen is desired
3
(Enabled) to enable it.
2
7-16
3
1
(Beep) if an audible
2