Chapter 7 Configuration and Setup
Alarms
To set a high, low, or STEL alarm level for the application, press
from the Set Parameters Menu. The appropriate Alarm Level Menu is displayed (see
Alarm Level menu, below). The last alarm level configured for that alarm is shown on
the display. If this is the alarm level for the present application, press
returns to the Application Setup Menu.
To change the alarm level, press
Level Enter menu, below). Key in the desired value and press
valid one, the display responds "-ACCEPTED-" and then return to the Application Setup
Menu.
Note: A Low alarm is an alarm activated by a reading above the lower of two alarm
setpoints; a high alarm is an alarm activated by a reading above the higher of the two
alarm setpoints. Either, or both, alarms may be selected.
STEL Time
To set a time interval for the application STEL alarm, press
Menu. The STEL Time Menu is displayed. The last time interval configured is shown
on the display. If this is the time interval for the present application, press
display returns to the Application Setup Menu. To change the time interval, press
STEL Time Select Menu is displayed (see STEL Time Select menu, below) Select the
time interval for the application (Off, 5, 10, or 15 minutes). The display responds "-
ACCEPTED-" and then return to the Application Setup Menu.
┌────────────────────────────────────────┐
│╒════METHANE SUMMARY════╕Application
││ HRL:
100.0ppm
││High:
100.0ppm
││ Low:
50.00ppm
││STEL:
5.00ppm @ 15min│
││TwaWt:
Off
││MolWt: 16.0
│└─────────Gas1──────────┘ =Prev
└────────────────────────────────────────┘
Alarm Level Menu
. The Alarm Level Enter Menu is displayed (see Alarm
1
┌────────────────────────────────────────┐
│╒════METHANE SUMMARY════╕Enter value for│
││ HRL:
100.0ppm
││High:
100.0ppm
││ Low:
50.00ppm
││STEL:
5.00ppm @ 15min│
││TwaWt:
Off
││MolWt: 16.0
│└─────────Gas1──────────┘ Enter=Accept
└────────────────────────────────────────┘
Alarm Level Enter Menu
│gas high alarm:│
500.0ppm
D: Off │
1=Change
│gas high alarm:│
25████ppm
(Zero=Off)
D: Off │
=Backspace
from the Set Parameters
5
7-24
,
, or
respectively
2
3
4
. The display
ESC
=Next│
. If the entry was a
ENTER
. The
ESC
. The
1