Menu Item
1 = Language
2 = Decimal
3 = Temp unit
4 = Press unit
5 = Conc unit
6 = Vol unit
GAS UNIT
Pressing 3 (Gas unit) from the Setup Menu causes the Unit Gas Selection Menu to be
displayed (see Unit Gas Selection menu, below). The current application is displayed on
the left and the menu of selections on the right. Choose the gas for which the units are
specified. (If there is a single gas application, this display is omitted and the display
shown in the right side of Gas Unit menu, shown below, is displayed instead.)
Note: Changing the gas unit from ppm to ppb does not increase the analyzer's
sensitivity.
Selections
Selects the language of the display: 1 = English, 2 = Francais, 3 =
Espanol, and 4 = Deutsche
Selects the type of decimal point character: 1 = a period (.) and 2 = a
comma (,).
Selects the temperature unit: 1 = °C and 2 = °F.
Selects the pressure unit: 1 = mmHg, 2 = inHg, 3 = kPa, 4 = bar, 5 =
psi, and 6 = A (atmospheres)
Selects the concentration default unit: 1 = ppb/ppm/% and 2 = mg/m 3 .
Selects the liquid volume unit: 1 = ml and 2 = cc.
┌────────────────────────────────────────┐
│╒══CURRENT APPLICATION══╕Select gas:
││ METHANE: ---- ppm
││
ETHANE: ---- ppm
││
BUTANE: ---- ppb
││ PROPANE: ---- ppm
││
HEXANE: ---- ppm
││GROUPTWA: ---- %
│└───────────────────────┘
└────────────────────────────────────────┘
Unit Gas Selection Menu
Chapter 7 Configuration and Setup
│1=METHANE
│2=ETHANE
│3=BUTANE
│4=PROPANE
│5=HEXANE
│6=GROUPTWA
Enter=All
7-5
│
│
│
│
│
│
│
│