Note: This table only contains P term values because this is all that is required for
humidity compensation.
Temperature Compensation
Temperature Compensation sets the temperature compensation off or on. The last
compensation condition configured is shown on the display. If this is the compensation
condition for the present application, press
parameter to be configured. Press
Temperature Compensation Select Display appears (see Temperature Compensation
Display, below). Select the condition for the application (Off or On). The display
responds "-ACCEPTED-" and then changes to the next parameter to be configured.
┌────────────────────────────────────────┐
│╒════════SUMMARY════════╕Application
││
││
││ Applic: METHANE
││
││
││
│└───────────────────────┘ =Prev
└────────────────────────────────────────┘
Pressure Compensation
Pressure Compensation sets the pressure compensation off or on. The last compensation
condition configured is shown on the display. If this is the compensation condition for
the present application, press
configured. Press
to change the compensation condition. The Pressure Compensation
1
Select Display appears (see Pressure Compensation Display, below). Select the condition
for the application (Off or On). The display responds "-ACCEPTED-" and then changes
to the Reference Compensation Summary.
Note: There is no pressure sensor in the instrument. If pressure compensation is
enabled the user is prompted to enter the ambient pressure prior to running any
application.
┌────────────────────────────────────────┐
│╒════════SUMMARY════════╕Application
││
││
││ Applic: METHANE
││
││
││
│└───────────────────────┘ =Prev
└────────────────────────────────────────┘
ENTER
to change the compensation condition. The
1
Temperature Compensation Display
. The display changes to the next parameter to be
ENTER
Pressure Compensation Display
7-32
Chapter 7 Configuration and Setup
. The display changes to the next
│ temp comp ref:│
Off
=Next│
│ pres comp ref:│
Off
=Next│