Thermo Scientific MIRAN 205B Series Instruction Manual page 184

Portable ambient air analyzer
Table of Contents

Advertisement

Appendix C Scan/Calibration Procedure
F.
CALIBRATION
1. Calculate five injection volumes that cover the complete concentration range
(also see Chapter 12). E.g., if the concentration is 0 to 1000 ppm, the
concentrations should be approximately 200, 400, 600, 800, and 1000 ppm.
For liquid injection volumes
For gas injection volumes
Where:
VI = injection volume in microliters (L)
C = concentration volume in parts per million (ppm)'
V
= total volume of cell and calibration system in liters (2.24L)
T
d =
Liquid density (g/cm
M =
Molecular weight of sample
P
Molar volu
R
Note : In the above example, 200 (the lowest concentration ) x 2.24 (the cell
volume) = an injection volume of 448 L. Since most syringes are not precise
enough to measure an injection volume of 448 L, select an injection volume
that can be accurately measured by the syringe. In the above example it will
be 450 L. Transpose the formula to C = V
concentration (C). In this example, C = 450/2.24 = 200.1. Record both the
concentrations and injection volumes.
2. From the Main Menu select "1=Analyze/Start"
3. Select "1=New Zero"
4. At the prompt "Install chemical filter", attach the Closed Loop Calibration
System and turn the pump on. Select "Enter=Continue"
5. Allow instrument to purge with "Zero Air" or clean air.. The instrument will
purge for 1 minute then proceed to set the Zero.
6. At the prompt for "Install Particulate filter", keep the Closed Loop Calibration
System attached. Press "Enter=Continue" and allow the instrument to purge
the Zero Air sample. When the purge is complete, select  or  until the
Absorbance screen appears. Verify that the absorbance equals zero(+/- .0010).
3
Depending upon the compound, either a gas or liquid may be injected,. Many compounds (e.g., benzene)
are not readily available in gaseous form, and are therefore inserted into the pump as a liquid. The pump
evaporates the liquid before it is introduced to the analyzer.
3
, use the formula:
3
, use the formula: VI = (C) (V
3
)
1
me
of
gas
at
25
24.4
C-10
 
V
T
VI
C
3
10
).
T
C
/V
, and solve for the
i
T
P
M
R
d

Advertisement

Table of Contents
loading

This manual is also suitable for:

Miran 205b sapphire

Table of Contents