Other Parameters
In the Continuous Sample Mode, the analyzer is placed in Auto Save and Auto Repeat
Mode. This means that the analyzer automatically stores the log without user
intervention and continue logging at the defined logging interval.
In Single Sample Mode, the Auto Save and Auto Report parameters are disabled. This
means that in order to save a log and then continue logging, it must be acknowledged.
The Auto Save and Auto Repeat in Single Sample Mode can be changed to the default
log options. To do this, press
enable Auto Repeat and 2 to enable Auto Save. When finished, press
selections. The display responds "-ACCEPTED-."
┌────────────────────────────────────────┐
│╒══════LOG SUMMARY══════╕
││
││Interval: 10 seconds
││
││
││Store:_OnAlrmOnly _STEL│
││√Mx √Min √Avg _Abs _Tmp│
│└───────────────────────┘
└────────────────────────────────────────┘
Alarms
To configure the state of the alarms, press
Page 1. The Alarm Menu is displayed (see Alarm menu, below). The current alarm
summary is displayed on the left and the menu of selections on the right. In the summary,
the alarms may be described as Off, Enabled, or Latched.
Note:
1. A Low alarm is activated by a reading above the lower of two alarm setpoints; a high
alarm is activated by a reading above the higher of the two alarm setpoints.
2. To configure the High, Low, and STEL alarm setpoints, see Alarms, below.
┌────────────────────────────────────────┐
│╒═════ALARM SUMMARY═════╕
││
││
││
││
││ √Conc Trig
││ √Avg Trig
│└───────────────────────┘6=Geiger countr│
└────────────────────────────────────────┘
(Other parms) from the Log Menu. Then press
5
Mode: Single sample│
Start: Manual
Other Parameters Menu
(Alarms) from the Analysis Setup Menu
2
High alarm: Enabled
Low alarm: Enabled
STEL alarm: Latched
√Relay 1 │4=Alarm trigger│
√Relay 2 │5=Relay output │
Alarm Menu
7-11
Chapter 7 Configuration and Setup
ENTER
_Auto repeat │
√Auto save
-ACCEPTED-
ALARM MENU
│1=High alarm
│2=Low alarm
│3=STEL alarm
─────
to
1
to set these