Chapter 7 Configuration and Setup
Wave Count
Wave Count sets the number of analytical wavelengths for the application. The last
number configured is shown on the display. If this is the number for the present
application, press
Press
to change the number. The Wave Count Enter Display appears. Key in the
1
desired number (1 to 10) and press
"-ACCEPTED-" and then changes to the Wavelength Summary.
Wavelength Summary
The Wavelength Summary reviews the wavelength position for each wave in the
application. The display also indicates whether all humidity compensation parameters
are zero or not (see Wavelength Summary Display, below). Move from one wave to the
next with the arrow keys. To accept this data for each wave, press
wave is reached, the display jumps to the Temperature Compensation Display. To
change the data in the Wavelength Summary for a wave, press
Position Display is presented.
Application Wavelength
Application Wavelength is the wavelength position defined for the installed application.
This parameter enables the position to be changed.
Note: If this is a standard library application, the value cannot be changed.
The last wavelength position configured is shown on the display. If this is correct, press
. The display changes to the next parameter to be configured. Press
ENTER
the wavelength position. The Wavelength Position Enter Display appears. Key in the
desired wavelength position (in m) and press
responds "- ACCEPTED-" and then changes to the next parameter to be configured.
Correction for humidity may be necessary as water vapor is a strong absorber in the
infrared, with frequently overlapping absorption bands of
. The display changes to the next parameter to be configured.
ENTER
ENTER
┌────────────────────────────────────────┐
│╒══WAVELENGTH SUMMARY═══╕
││
││
: 3.333um
││
││ _Humidity compensation│
││
parameters installed │
││
│└─────────Wave1─────────┘ =Prev
└────────────────────────────────────────┘
Wavelength Summary Display
. If the entry is valid, the display responds
1
─────
Wavelength
Summary
─────
1=Change
=Next│
. If the entry is valid, the display
ENTER
7-30
. When the last
ENTER
. The Wavelength
to change
1