Appendix G Using Hyperterminal with the MIRAN SapphIRe
Note: The SapphIRe baud rate is user-selectable. While 19,200 baud is optimum, it
may not be correct. The terminal program should be set to match the analyzer under
5=Report/Data, 3=Data transfer, 2=Change baud.
The analyzer must be configured for TEXT-MANUAL transfer protocol before using a
rd
3
party terminal application. From the main menu, this is done as follows:
5=Report/Data, 3=Data transfer, 1=Chnge mode, 2=Manual.
Once the SapphIRe is set up for TEXT-MANUAL and the terminal program is running,
data may be sent from the analyzer to the PC. First, the terminal program should be set to
write all incoming data to a file (capture the data). This is done in Hyperterminal by
selecting the menu item Transfer... Capture Text, then setting the file name. Next, the
transfer is originated at the SapphIRe by selecting 3=Send, then selecting the appropriate
data from the SEND MENU: 1=Site, 2=Full setup, 3=Partial setup, 4=Library names.
To receive data on the SapphIRe from the PC, choose 4=Receive from the DATA
TRANSFER menu on the SapphIRe then use the menu item Transfer... Send Text in
Hyperterminal.
The type of data is included inside each of the text files, so the file type does not have to
be selected before receiving data on the SapphIRe.