Thermo Scientific MIRAN 205B Series Instruction Manual page 72

Portable ambient air analyzer
Table of Contents

Advertisement

Chapter 7 Configuration and Setup
If 1 is pressed, the next display requires entering the zero warning time interval. This can
be set from 0 to 999 minutes or hours. The numeric field entry is keyed in; the units are
selected with the down arrow key. The display responds "-ACCEPTED-."
If 2 is pressed, the next display requires entering the reference absorbance drift. This can
be a number from 0 to 9.9999. The display responds "-ACCEPTED-." A typical
deviation is 0.015 absorbance units.
Note: If drift level is not zero and the reference wavelength is not measured, there is a
warning that the reference wavelength must be measured.
Autozero
This feature is not available at the present time.
View Enable
The View Enable parameter enables specifying which types of analysis data may be
viewed. To access this parameter, press
Page 2. Upon presentation of the next display, press
Concentration,
2
Intensity,
for Intensity gain,
7
press
to set these selections. The display responds "-ACCEPTED-."
ENTER
In the summary, data views are marked active (√) as follows:
Conc:
Concentration view enabled
Max:
Concentration maximum view enabled
Min:
Concentration minimum view enabled
Avg:
Concentration average view enabled
STEL:
Concentration STEL view enabled
Absb:
Absorbance view enabled
Intn:
Intensity view enabled
Gain:
Intensity gain view enabled
STEL Max:
Concentration STEL maximum view enabled*
Time:
Maximum, Minimum, and/or STEL Maximum time views enabled*
*STEL Max and Time data views are not shown in the above screen. However, they are
displayed when View Enable is accessed.
Note: Absorbance, Intensity, and Intensity Gain views are primarily used by factory
personnel.
for Maximum,
for Minimum,
3
for STEL, and
8
(View Enable) from the Analysis Setup Menu
4
to specify time,
0
for Average,
4
for STEL maximum. When finished,
9
7-18
for
1
for Absorbance,
for
5
6

Advertisement

Table of Contents
loading

This manual is also suitable for:

Miran 205b sapphire

Table of Contents