Built-in clock's date/time is
wrong!
CHECK PRINTER!
Log data memory invalid!
All sites cleared. Probable
cause is turning power off
during memory access.
Pathlength setting is wrong.
RECOMMENDED
SERVICE IS DUE!
Spectrometer startup has
failed! Try powering off,
then back on.
Watchdog timer has reset
the system!
Table 4-4. Warning Messages (continued)
This message is only generated during the
initial self-test on powerup. It indicates that
the time reported by the real time clock
hardware is invalid. When this message is
displayed, the clock is set to a valid time. If
the analyzer power is cycled again and this
message still appears, the backup battery
may be dead.
This message appears if a printer attached to
the printer port is reporting an error
condition such as: out of paper, offline, or
error.
This message indicates that the analyzer
found a problem in the data stored in
logging memory. This could be caused by
the analyzer's power being interrupted
during a memory write or by low backup
battery. If this message appears, all logged
data is deleted. Try logging the data again.
This warning message appears if the
pathlength setting is changed to an invalid
position during analysis.
This message appears when the analyzer's
next service date has passed. Annual
maintenance reduces field breakdowns.
Note that normal operation is possible
should this message appear.
This message is displayed only during
power-on self-test if communications
cannot be started with the spectrometer
controller board. As the message states, try
powering off the analyzer and then turn it
on again.
This warning message is displayed only
during the power-on self-test if the system
was reset by the watchdog. As the message
indicates, this error is most likely caused by
a software problem.
4-15
Chapter 4 Operation
Contact service for
assistance.
Clear the status at the
printer.
Ensure the analyzer is
properly powered. If
the problem persists,
call service for
assistance.
Set the pathlength
changer back to the
correct position and
restart analysis.
Contact service for a
service update.
Cycle power off/on. If
the message persists,
return the analyzer for
service.
Note what the analyzer
was doing before it shut
off, and contact service
for assistance