Chapter 11 Firmware Update
In the Firmware Main Menu, press
In the Serial Menu, press
ACCEPTED" and returns to the Serial Menu. Press
Menu. Then press
firmware is about to be erase. Press
The Loading Firmware screen is displayed. In ThermoConnect, highlight the M205.prm
file and press
F6
minutes.
When complete, the Firmware Main Menu is displayed (see Firmware Main Menu,
above). Verify that firmware revision is correct and validity states are valid. Press
exit. The analyzer restarts at this point and error messages related to the application are
displayed. Proceed to the next section to reload the specific application.
╔════════════════════════════════════════╗
║───────────────MAIN MENU────────────────║
Firmware version: 205
Firmware validity: Valid
1=Update firmware
2=Validate firmware
4=Details
╚════════════════════════════════════════╝
Firmware Main Menu
to display the Serial Menu.
3
to select 19200 baud. The display briefly read "19200
1
or
to erase the present firmware. The display warns that the
ENTER
1
ENTER
╔════════════════════════════════════════╗
║──────────────SERIAL MENU───────────────║
1=Verify serial link
╚════════════════════════════════════════╝
to send the new software to the analyzer. This process takes about 25
╔════════════════════════════════════════╗
║────────────LOADING FIRMWARE────────────║
Loading firmware using serial port... ║
Waiting...
╚════════════════════════════════════════╝
Loading Firmware Screen
050000P
3=Serial setup
5=Exit update mode ║
to return to the Firmware Main
ESC
to continue.
Baud Rate
19200
2=Baud rate
Serial Menu
Esc=Cancel ║
11-4
to
5