Torque Output Signal, Code Kf1; Torque Output Signal, Code Sf1/Su2; Setting Up Zero Point - HBM T10F KF1 Operating Manual

Torque flange
Table of Contents

Advertisement

34
SUNSTAR传感与控制 http://www.sensor-ic.com/ TEL:0755-83376549 FAX:0755-83376182 E-MAIL:szss20@163.com

7.1 Torque output signal, code KF1

The factory setting for the frequency output voltage is 12 V (asymmetrical).
The frequency signal is on pin 4 opposite pin 1. No voltage switching is
possible.

7.2 Torque output signal, code SF1/SU2

The factory setting for the frequency output voltage is 5 V (symmetric, RS 422
complementary signals). The frequency signal is on pin 4 opposite pin 1. You
may switch the output voltage to 12 V (asymmetrical). To do this you must set
switches S1 and S2 to position 1 (in this case pin 1 →
Fig. 7.2: Switches for changing the frequency output voltage

7.3 Setting up zero point

If you have a torque flange with the voltage output option (SU2), you can
remove the identification plate to gain access to two potentiometers. You can
use the zero point potentiometer to correct certain zero point variances. The
range of compensation is minimum "400 mV at nominal amplification. The
end point potentiometer is used for balancing at the factory and is capped with
varnish so that it cannot be turned unintentionally.