7-3-3-4 Zero & Span All Calibration
Control..
Adv.Calibration..
Cancel calibration!
ZeroAll!
SpanAll!
ZeroSpanAll!
7-18
7-3-3
Advanced Calibration
Before selecting any further line
make sure the required calibra-
tion gas is applied!
Supply all calibration gases with
the same flow as the sample gas
(recommeded approx. 1 l/min),
pressureless and utilizing the
right gas fitting (see sect. 3-4).
Make sure the calibration purge
time is set to a value ensuring the
measuring cell is filled pro-perly
with the related calibration gas
after the valve has opened!
Ensure the warm-up time after
switching on has elapsed!
Warm-up time is 15 to 50 minutes
depending on installed measu-
ring system and configuration!
Starting from the measurement screen press
the DOWN key to open the MAIN MENU and
enter the CONTROL - ADV.CALIBRATION
menu.
To start a ZERO & SPAN calibration for ALL
channels select the last line.
Note!
Single channel analyzers show the same
menu, with the restriction, that the term "ALL"
relates to the single channel only!
The analyzer immediately begins zero
calibration(s for all channels), followed by span
calibration(s). Watch the screen for status
information (explained by means of exemplary
pictures on the next page).