Tools And Test Equipment; General Troubleshooting And Diagnostics Information; Diagnostics Menu - Emerson Rosemount CT5100 Reference Manual

Continuous gas analyzer
Hide thumbs Also See for Rosemount CT5100:
Table of Contents

Advertisement

Reference Manual
00809-0100-4512
11.5

Tools and test equipment

All tools required when performing failure diagnosis are standard hand tools.
The only test equipment required when performing failure diagnosis is a multi-meter that
may be used to perform continuity checks on electrical wiring during an inspection.
11.6
General troubleshooting and diagnostics
information
The Rosemount
automatically resolve system issues, and to recover from power failures and return to a
normal working state. This troubleshooting guide is intended to assist maintenance
personnel when the analyzer has not appeared to be working normally for a period of
more than five minutes. If the procedures given in this section fail to return the analyzer to
normal operation or do not identify a fault, notify your service agent for further assistance.
11.7

Diagnostics menu

Selecting DIAGNOSTICS on the Main Menu will open the Diagnostics Menu
The Diagnostics Menu lists those main components of the Rosemount CT5100(Ex) where
problems can be diagnosed using the BIST system. The Diagnostics Menu also enables you
to check on the status and, where appropriate, the values of, any of the listed
components.
Press the REFR button (F) to refresh the fault list.
Use the Scroll Up (C) and Scroll Down (G) buttons to highlight a component for further
investigation by the BIST system.
Once you have highlighted the component you wish to diagnose, press the SEL button (B)
to select that fault. The Diagnostics screen for that component will then open.
Reference Manual
CT5100(Ex) is specifically designed to run unattended indefinitely, to
Troubleshooting and diagnostics
October 2019
(Figure
11-8).
119