Advertisement

4.10.5. Setting the DO output

4.10.5.1. How to validate outputting the FULL SCALE 2

Description
 Select a contact output as DO1 and/or DO2 at FULL SCALE2 measurement status.
For actual keying, refer to the typical operation indicated below. Set the protection to OFF beforehand. (See Section 4.4.1.)
Operation
Set the DO1 output to "FULL SCALE2".
(example)
Also, set the contact to "ACTIVE ON".
Key operation
Press the
Press the
ENT
Press the
* Press the
Press the
ENT
Press the
Press the
ENT
Press the
ENT
Press the
ENT
* To select normally on, press the
ESC
Press the
the measurement mode.
- 52 -
Description
key twice to display "OUTPUT SETUP".
key once to display "ZERO ADJUSTMENT".
ENT
key for 6 times to display "DO1 OUT".
key again to display "DO2 OUT".
key once to blink the cursor.
ENT
key for 3 times to display "FULL SCALE2" on the 2nd line.
key once to register "FULL SCALE2".
ENT
――― "FULL SCALE2" has been registered. ―――
key once to display "CONTACT ACTION".
ENT
key once to register "ACTIVE ON"(normally off).
ENT
――― ACTIVE ON has been registered. ―――
ESC
key twice and then press the
key.
key for 3 times to enter
Display
OUTPUT SETUP
ZERO ADJUSTMENT
SET ZERO
DO1 OUT
NOT USED
DO1 OUT
NOT USED
DO1 OUT
FULL SCALE2
DO1 OUT
 COMPLETE 
STATUS OUT
CONTACT ACTION
CONTACT ACTION
ACTIVE ON
CONTACT ACTION
 COMPLETE 
STATUS OUT
CONTACT ACTION
0.000
%
0.000
m3
INF-TN3FLR-E