Auto Calibration - Fuji Electric ZAF-6 Instruction Manual

Thermal conductivity gas analyzer thermomat
Table of Contents

Advertisement

6.3 Setting of auto calibration
(When auto calibration contact output has been selected)

6.3.1 Auto calibration

(1) During measurement, press the
Calibration" by pressing the
key. Press the
(2)
make the setting by pressing the
the
key, and then press the
Make the setting for each item by using
the
,
the
key.
ENT
To close Setting of Auto items
To close the "Setting of Auto calibration" or
cancel this mode midway, press the
A previous screen will return.
Description of setting items
Setting range
Cycle
: 1 to 99 hours or 1 to 40 days (initial value 7days)
Flow time
: 60 to 599 sec
30
key to
MODE
or the
key.
ENT
key.
ENT
or the
key, and then press
key.
ESC
User Mode
Setting about Calibration
Alarm Setting
Setting of Auto Calibration
Changeover of Range
Parameter Setting
About Auto Cal.
or
Start Time
Cycle
Flow Time
ON/OFF
Stop Auto Calibration
(initial value 300sec)
Select an item
with UP/DOWN and ENT
Select setting item
SUN 12 : 00
07
day
300 sec.
OFF
Time : MON 17 : 24
INZ-TN6ZAF-E