Operation Vss500N Generator - Ametek burst NX8 Quick Start Manual And Safety Instructions

Surge, burst, damped oscillatory, telecom surge generators
Table of Contents

Advertisement

AMETEK CTS
1
SYNC input
2
Reference earth connection
3
Connection warning lamp
4
Ventilator
5
Safety circuit
6
External trigger
7
Power on switch, fuses
10
Parallel interface GPIB / IEEE 488
IEEE 488 interface with IEEE connector
11
Opto Link Serial interface (USB)
- Optical USB interface "USB" connector.
For data transfer an optical USB interface is available. The internal RS 232 interface is converted to USB standard.
Therefore the user must set the same Baudrate in the device and control software.
For remote control the opto link must be used.
12
Remote control connector CN
External coupling devices are controlled via this remote control connector.
13
Fail detection FAIL 1 EUT control (TEST STOP)
Grounding this input will cause a complete stop of the running test procedure. (+15V to ground) The test must be completely
restarted.
Fail detection FAIL 2 EUT control (TEST PAUSE)
Grounding this input will cause a break for the running test procedure (+15V to ground).The test will be continued when the
input is no more connected to ground.
8.4.

Operation VSS500N generator

The simulator VSS500N12.x is operated by an easy menu control system. Seven
function keys are available to select parameters and functions. All functions are
indicated on the display.
The selected parameter is blinking and can be changed by turning the knob (incr./decr.). The takeover
of the input value occurs after about 500ms encoder downtime. This allows the operator a brief check
of the correct input value.
➔ :
ESC :
EXIT :
The digit to be changed can be selected with the cursor (➔ ).
- Setted values are direct indicated on the screen.
- Status on the bottom lines shows thedesired status after pressing the function key.
ESC will take you back to the previous level in the menu and set the displayed values. The
latest settings are stored automatically and will be recalled when the menu is selected again.
The firmware will reset to the main screen.
1.00
Quick start guide - Single pulse generators
8
Mains selector 115V / 230V
9
Fuse of the high voltage power supply
10
Parallel interface GPIB / IEEE 488
11
Opto Link interface (USB)
12
Remote control connector CN
13
FAIL 1 / 2 Input
53 / 88