Technical Data - Slow Damped Oscillatory As Per Iec 61000-4-18; Technical Data - Fast Damped Oscillatory As Per Iec 61000-4-18; Technical Data - Ring Wave As Per Iec 61000-4-12 - Ametek burst NX8 Quick Start Manual And Safety Instructions

Surge, burst, damped oscillatory, telecom surge generators
Table of Contents

Advertisement

AMETEK CTS
10.11.

Technical data - Slow damped oscillatory as per IEC 61000-4-18

Test Level
Output voltage
Wave form (open circuit):
Rise time
Oscillation frequency
Decaying
Source impedance
Coupling capacitors
Polarity
Repetition rate
10.12.

Technical data - Fast damped oscillatory as per IEC 61000-4-18

Test Level
Output voltage
Wave form (open circuit):
Rise time
Oscillation frequency
Decaying (voltage)
Source impedance
Coupling capacitor
Polarity
Repetition rate
Burst duration, td
Burst period
Short circuit current
Rise time current
Decaying (current)
10.13.

Technical data - Ring Wave as per IEC 61000-4-12

Test level
Open circuit voltage
Wave shape open circuit
Rise time first peak T1
Oscillation frequency 1/T
Decaying of Pk1 to Pk2
Decaying of Pk2 to Pk3
Decaying of Pk3 to Pk4
Decaying other peaks
Output impedance
Wave shape short circuit
Rise time first peak tr T1
Oscillation frequency 1/T
Decaying
Short circuit current 12Ω
Short circuit current 30Ω
Polarity
Repetition rate
Events preselection
Counter
250 V – 2`500 V  10 %
250 V – 3`000 V  10 %
75 ns  20 %
100 kHz and 1 MHz  10 %
Peak 5 must be > 50 % of peak 1 value
Peak 10 must be < 50 % of peak 1 value
No requirements for other peaks
200   20 %
0,5 µF  10 %
Positive, Negative
At least 40/ s for 100 kHz and
At least 400/ s for 1 MHz
450V – 4`400 V  10 %
5 ns  30 %
3 MHz, 10 MHz, and 30 MHz  10 %
Peak 5 must be > 50 % of peak 1 value
Peak 10 must be < 50 % of peak 1 value
No requirements for other peaks
50   20 %
33 nF  10 %
Positive, Negative
5000 /s
50 ms  20 %
at 3 MHz
15 ms  20 %
at 10 MHz
5 ms  20 %
at 30 MHz
300 ms  20 %
9 A to 88 A  20 %
< 330 ns
at 3 MHz
< 100 ns
at 10 MHz
< 33 ns
at 30 MHz
Peak 5 must be > 25 % of peak 1 value
Peak 10 must be < 25 % of peak 1 value
No requirements for other peaks
250 V – 6'000 V ± 10 %
0.5 μs ± 30 %
100 kHz ± 10 %
40 % - 110 %
40 % - 80 %
40 % - 80 %
no requirements for other peaks
12 Ω, 30 Ω
± 20 % (internal depend setting)
12 Ω internal ± 20 % (with setting R external)
0.2 to 1.0 s (short circuit)
100 kHz ± 10 %
no requirements
500 A ± 10 %
200 A ± 10 %
(+10 % - 0 %, models as per IEC 61008-1 for RCCB)
positive / negative / alternating
250 V – 4000 V
1 s – 999 s
> 4000V 10 s – 999 s
1 - 30'000 or endless
1 – 100'0000
1.00
Quick start guide - Single pulse generators
HV out on model N6.5; N6.6; N6.7; N6.8
at
77 / 88