9.3.

Front view

1
Display
2
"Test On"
Function keys "F1..F7"
3
4
Knob (Inc / Dec)
5
Cursor keys "" and "→"
1
Display
All functions and parameters are displayed (8 lines with max. 40 characters).
2
Test On
By pressing the key "TEST ON" the test procedure is initiated with the preselected parameters. The yellow key is illuminated
in TEST ON mode.
3
Function keys "F1 .. F7
Parameters and functions, displayed in the lowest line, can be selected with the related function key.
4
Knob ( Inc / Dec )
The knob increments or decrements test parameters with a numeric value or selects from a list of parameters.
5
Cursor keys
Parameters and functions can be changed on-line. The selection of these parameters is realized with the cursor moving to
the left or to the right.
6
Exit
Pressing of the Exit function will cause a reset of the firmware. This is only possible if no test routine is running.
7
ESC
When pressing the ESC button the user moves back one page in the menu.
8
BNC CRO U
At the BNC output the voltage pulse of the generator can be measured.
Dividing ratio is 10:1 for all pulses. Only for measuring residual voltage across VDR's!
9
BNC CRO I
At the BNC output the current pulse of the generator can be measured.
Dividing ratio depends on the selected pulse. 1:1 (18A), 10:1 (140A)
10
BNC - CRO Trigger
At the BNC output the generator trigger pulse is available (+5 V rectangular). This output can be generally used as
oscilloscope trigger output and is synchronous to the surge impulse
11/12
8/20s current pulse output.
The 8/20s current pulse can be used under direct short circuit conditions between the L and PE output plugs. (Front panel
use). This pulse is also available in the special test box arrangement which can be plugged into the cover panel of the
generator.
HV Output on top cover of CSS 500 N2
6
Exit
7
Escape
8
CRO U ( surge)
9
CRO I ( surge)
1.00
Quick start guide - Single pulse generators
CRO trigger output  5V
10
11
8/20s current pulse high
12
8/20s current pulse low
and 100:1 (1200A).
58 / 88