Ametek burst NX8 Quick Start Manual And Safety Instructions page 40

Surge, burst, damped oscillatory, telecom surge generators
Table of Contents

Advertisement

AMETEK CTS
1
Reference earth connection
2
Ventilator
3
Power On Switch
4
Mains input
5
External Trigger IN
6
SYNC input (External)
External synchronization input for synchronize the generator impulse to this signal.
Input signal range: 10 V – 690 V.15Hz to 500 Hz
For each test with synchronization the generator measures the frequency. If no signal is detected, a window will propose to
continue in asynchronous mode.
7
Sys Link OUT
Sys link is the internal control bus to the connected vsurge NX devices. The sys link is a daisy chain wired bus for connect
the devices in series. The bus includes a 26 pole high density connectors. This port is terminated with the SWL AD (Safety
Warning Lamp Adapter).
8
Ethernet interface
Ethernet interface for remote control from an external computer.
9
Opto Link interface
Optical interface for remote control from an external computer USB interface.
10
Fail input EUT 1, EUT 2
Programmable input monitors for fail events. Possible input settings: Notify, Stop, Break or Disabled
Input signal:
Max- Input voltage:
6
Sync input
7
Sys Link OUT
8
Ethernet interface
9
Opto Link Interface
10 Fail, EUT 1, EUT2
5 V pos. slope
+24 V
Quick start guide - Single pulse generators
1.00
11 Sys Link IN
12 USB B interface
13 USB A interface
40 / 88