Maintenance, Adjustments, Replacement Of Parts; Disposal Of Devices; Calibration And Verification - Ametek burst NX8 Quick Start Manual And Safety Instructions

Surge, burst, damped oscillatory, telecom surge generators
Table of Contents

Advertisement

AMETEK CTS
13.

Maintenance, Adjustments, Replacement of Parts

Electrical maintenance must only be performed by qualified service technicians
CAUTION
The generators do not contain any parts or components requiring special maintenance.
Electrical maintenance must only be performed by experienced and specially trained technicians. Generally, standard
maintenance requires only the periodic cleaning of the instrument, verification and calibration of certain parameters.
When removing the cover or other parts of the equipment, high voltage parts may become exposed. High voltages are
potentially lethal.
For service, repair, adjustment or replacement of parts, the generator must be disconnected from all power supply
sources before covers are to be removed.
The user is not permitted to change or modify any EM TEST / TESEQ generator. Only original EM TEST / TESEQ parts
and components shall be used for repair and service. AMETEK CTS is not responsible for accidents or injuries caused
through the use of parts or components not sold by AMETEK CTS.
Maintenance and service must only be performed by qualified service technicians who are trained and familiar with the
dangers of servicing the EM TEST / TESEQ generator.
Only fuses of correct voltage and amperage as specified by the manufacturer are to be used for replacement. The repair
of fuses is not permitted.
13.1.

Disposal of devices

For the disposal of electronic devices, the country-specific regulations are to be considered. The equipment
should be delivered to a specialized waste collection center.
EM TEST and Teseq devices can be returned to AMETEK CTS in Switzerland or to their agency for
adequate disposal. Alternatively, the equipment can be handed over to a specialized enterprise for disposal
of electronic devices.
Details to used material and components
- Built in capacitors contain no polychlorinated biphenyl (PCB).
- Back-up batteries and rechargeable batteries must be disposed separately.
13.2.

Calibration and Verification

For the calibration and verification of the voltage impulses the using of calibration adapters is recommended.
The calibration is at all times without any mains supply on the output to calibrate. Therefore, the following warnings are to
adhere strictly.
Before starting the calibration or verification remove the EUT Mains Supply from the generator and from
the coupling network
CAUTION
Quick start guide - Single pulse generators
1.00
88 / 88