Rear View - Ametek burst NX8 Quick Start Manual And Safety Instructions

Surge, burst, damped oscillatory, telecom surge generators
Table of Contents

Advertisement

AMETEK CTS
8.3.

Rear View

1
SYNC input
2 Reference earth connection
3 Connection warning lamp
4 Ventilator
5 Safety circuit
6
External trigger
7
Power on switch, fuses
1
SYNC input
The release of single surge pulses can be synchronized to the signal at the sync input. Normally this input may be connected
directly to the phase of the power mains supply. If no signal is connected the generator automatically works in asynchronous
mode.
2
Reference earth connection
The generator has to be connected to the reference earth plane of the test set up.
3
Warning lamp
A voltage free contact(230V 6A) is available for external warning indications (warning lamp). The signal is generated after
pressing TEST ON.
4
Ventilation
After long term duration tests the generator should keep on running for some minutes to cool down the system.
5
Safety circuit
The test can only be started if the security circuit is closed. If the circuit is opened during a running test the simulator will be
switched off immediately.
6
External trigger
One single transient event can be released. Trigger level 5-15V positive going.
7
Power on switch
The switch is part of the mains filter. Mains fuses are part of the filter. (230V / 2x 1A and 115V / 2x 2A )
8
Mains selector
Selection of 115V / 230V
9
Fuse of the high voltage power supply
The high voltage power supply is protected by this fuse „F3". In case that no high voltage is generated but the control unit
works properly this fuse shall be checked.
Quick start guide - Single pulse generators
8
Mains selector 115V / 230V
9
Fuse of the high voltage power supply
10
Parallel interface GPIB / IEEE 488
11
Opto Link interface (USB)
12
Remote control connector CN
13
FAIL 1 / 2 Input
1.00
52 / 88