Safety Circuit, Warning Lamps - Ametek burst NX8 Quick Start Manual And Safety Instructions

Surge, burst, damped oscillatory, telecom surge generators
Table of Contents

Advertisement

AMETEK CTS
7.8.

Safety circuit, Warning lamps

Safety circuit and warning lamps are located at each end of the Sys Link. Each End of the Sys.Link is terminated with an
adapter.
Safety circuit
Plug position:
SYS.LINK IN vsurge NX
generator
Connector:
2 screw contacts IN, OUT
Pin Sys Link:
7, 16
Operating:
short circuit closed =
operation
short circuit open = no
operation
Voltage level:
24 VDC (internal powered)
Warning Lamp
Plug position:
SYS.LINK OUT Last device
from daisy chain
Connector:
2 screw contacts IN, OUT
Pin Sys Link:
8, 17
Operating:
Potential free contact
open = System is out of
operation (green)
closed = System is operating
(red)
Voltage level:
60 V, max. 2 A (external
powered)
Important for operation
Connect the delivered Safety Circuit Terminal STC or Safety
Circuit Adapter SCC AD to the SYSLINK plug.
The vsurge NX15 generator does not start any test if the safety
circuit is not connected and closed.
NOTE: The short circuit is already short-circuited at the delivered
IMPORTANT
Safety circuit adapter SCC AD.
Quick start guide - Single pulse generators
1.00
35 / 88