11.10. Technical Data - Ametek burst NX8 Quick Start Manual And Safety Instructions

Surge, burst, damped oscillatory, telecom surge generators
Table of Contents

Advertisement

AMETEK CTS
The user is responsible to control the rating and the integrity of all cables connected to the NSG
3150 generator, especially cables connected to the Surge output.

11.10. Technical data

Parameter
Pulse voltage (open circuit):
Pulse current (short circuit):
Impedance:
Polarity:
Phase synchronization:
Coupling:
Pulse repetition:
Test duration:
Instrument supply
Fuse
Internal disruption of the electronics can result in the interference voltage or the EUT
supply voltage being present on the EUT's outer casing.
Electrical breakdown or arcing from connections that are overstressed voltage wise
during the test.
Explosion of components with fire or fragmentation as a result of energy dissipated,
e.g. from the resultant supply current or ignition of vaporized plastic materials.
Faulty behavior by the EUT, e.g. a robot arm strikes out or a temperature controller fails,
etc.
Value
± 500 V to 15.0 kV (in 1 V steps)
± 250 A to 7.5 kA
2 / 12 Ω
Positive / negative / alternate
Asynchronous, synchronous 0° to 359º (in 1º steps)
IEC / external / manual
10* .... 600 s (in 1 sec steps) 1 .... 10 min.
500 V
4401 V
8001 V
12001 V
1 to 9999 pulses, Continuous
85 to 265 VAC, 50 / 60 Hz (max. 3.15 A)
3.15 A slow blow
Quick start guide - Single pulse generators
– 4400 V
10 s @ positive, negative
20 s alternate
– 8000 V
20 s
– 12000 V
35 s
– 15000 V
50 s
1.00
82 / 88