Quick Start - Ametek burst NX8 Quick Start Manual And Safety Instructions

Surge, burst, damped oscillatory, telecom surge generators
Table of Contents

Advertisement

AMETEK CTS
Page 2
CSURGE (1200A)
F1 : Quick Start
F2 : Usual Routine
F3 : Change Voltage after n by V
F4 : Change Phase angle after n by A
F5 : Change Polarity after n pulses
F1
F2
F3
F1 Quick Start
Easy and fast operation of the generator. All parameters can be adjusted during a running test routine.
F2 Usual Routine
Usual routine for simple tests without any iteration. All parameters can be adjusted and stored.
F3 Change charging voltage after n pulses by V
The charging voltage V is changed from V1 to V2. After the preselected number pulses the test level is changed by V until V2
is reached.
The same parameters as under Quick Start are selectable. For the limitation of the max. admissible repetition rate the higher
value of V1 and V2 is valid.
F4 Change the phase angle after n pulses by A
The phase angle related to which the surge current pulse is released is changed from A1 to A2. After the preselected number
of n pulses the actual phase angle is changed by A until A2 is reached. The same parameters as under Quick Start are
selectable.
F5 Change polarity after n pulses
After the release of the preselected number of pulses the polarity is changed. The procedure always starts with positive polarity
and changes than to negative. The same parameters as under Quick Start are selectable.
9.7.

Quick Start

Easy and very fast operation of all standard functions of the equipment. The latest simulator settings are stored automatically
and will be recalled when Quick Start is next selected.
CSURGE (1200A) QUICK START
Vch
= 2000V
+/-
= +
tri
= Auto
START CHANGE
F1
F2
F3
Press START and the test routines begin to work.
Press CHANGE and the actual parameters can be changed.
All function keys except F2 (manual trigger) can Stop the test routine.
Page 3 (Start)
CSURGE (1200A)
QUICK START
Vch
= 2000V
+/-
= +
tri
= Auto
Vch = 2000 V
U =
STOP
I =
F1
F2
F3
The user can select the parameter with the related function key and can change the value with the knob. The cursor allows
the user to define the value of the digit which should be changed (fast or slow change).
Pressing the ESC button will bring the user back to the previous level from where the test can be restarted with new levels.
After restart the actual test time is displayed. All functions keys except F2 (MAN TRIGGER) can stop the test routine. The
latest setting will be displayed.
F4
F5
F6
A
=
0 dgr
rep
=
10s
n
=
endl
F4
F5
F6
A
=
0 dgr
rep
=
10s
n
=
endl
+ 30V
COUNTER
+ 1000A
0000043
F4
F5
F6
Quick start guide - Single pulse generators
F7
F7
F7
1.00
62 / 88