AMETEK CTS
7.10.

Rear view vsurge NX15

1
Reference earth connection
2
Ventilator
3
Power On Switch
4
Mains input
5
External Trigger IN
1
Reference earth connection
The generator must be connected to the reference earth plane of the test set up. The connection at the rear part of the
generator is an alternative to the grounding point at the front panel
2
Ventilation
A power controlled ventilator is cooling the vsurge NX15 generator. During the most application, the ventilator runs with
variable speed depends the cooling requirement.
After long term duration tests the generator should keep on running for some minutes to cool down the system.
3
Power on switch
The switch is part of the mains filter. Mains fuses are part of the filter. (230V / 2A and 115V / 4A)
4
Mains Input and Fuse
Mains input for the generator power supply.
85 V – 254 V ac
Input voltage:
Fuses:
115 V:
230 V:
5
External trigger IN
One single event, burst, surge, voltage dip can be released. Trigger level 5V positive going.
BNC connector
Digital Input
Positive slope
Trigger Level:
Max- Input voltage: +24 V
6
Sync input
7
Sys Link OUT
8
Ethernet interface
9
Opto Link Interface
10 Fail, EUT 1, EUT2
4 A slow blow, 5 x 20 mm.
2 A slow blow, 5 x 20 mm
50 Ω
(0 – 5 V)
2.5 V +/-1 V
Quick start guide - Single pulse generators
1.00
11 Sys Link IN
12 USB B interface
13 USB A interface
39 / 88