Ametek burst NX8 Quick Start Manual And Safety Instructions page 38

Surge, burst, damped oscillatory, telecom surge generators
Table of Contents

Advertisement

AMETEK CTS
1
Active indication
2
USB input (Memory stick)
3
Touch screen
4
Back
5
Knob (Inc. / Dec /Enter)
7
START / PAUSE / STOP button
- Green indication: Start or continues a test
- Orange indication: Pause of a running test
- Red indication: Stop a running or paused test.
8
CRO I (surge)
At the BNC output, the current pulse (surge) of the generator can be measured galvanic separated. This output can be used
for get a fast overview. For accurate measurements please use external measuring devices.
- Impedance measuring instrument: >1MΩ.
- Max. output voltage:
9
CRO V (surge)
At the BNC output the voltage pulse (surge) of the generator can be measured galvanic separated. This output can be used
for get a fast overview. For accurate measurements please use external measuring devices.
- Impedance measuring instrument: >1MΩ.
- Max. output voltage: 10V 10% at 15 kV impulse voltage.
10
BNC - CRO Trigger
At the BNC output, the generator trigger can be checked for trigger a scope, (+5 V rectangular).
11
LED bar
Indicates the charging of the internal impulse capacitor.
12
High/Low-output
Pulse output 500 – 15.000 V, impedance 40, 50 and 500 Ohm (selectable).
Connector: HIVOLT HC52RB (appropriate for max. 20 kV)
6
" Test On"
7
Start / Pause / Stop
8
CRO I (surge)
9
CRO V (surge)
CRO Trigger output  5V
10
10V 10% at 375 A output current (Z = 40 Ohm)
10V 10% at 300 A output current (Z = 50 Ohm)
10V 10% at 30 A output current (Z = 500 Ohm)
1.00
Quick start guide - Single pulse generators
11
LED bar for charging voltage
12
High/Low-output
38 / 88